1.
Chinese Medical Journal ; (24): 761-766, 2007.
Article in English | WPRIM | ID: wpr-240334

ABSTRACT

<p><b>BACKGROUND</b>Percutaneous kyphoplasty (PKP) using balloon expander has been proved to be effective in the treatment of painful vertebral compression fractures. Recently, Sky-bone expander, an alternative bone expander for PKP has been developed. The purpose of this study was to show our preliminary clinical outcomes of PKP with Sky-bone expander.</p><p><b>METHODS</b>PKP with Sky-bone expander was performed in 25 patients (30 vertebrae). The operation time, bleeding volume, cement volume injected were recorded. The pain and functional activities of the patients before and after the operation were compared using Wilcoxon signed-rank test. The cement distribution in the vertebrae, vertebral height restoration, and kyphosis correction after the procedure were evaluated by radiography. The pre- and post-operative absolute values of the vertebral height and kyphotic angle were compared by paired-sample t test. All the patients were followed up by telephone or clinic consulting after being discharged from our hospital.</p><p><b>RESULTS</b>The procedure was performed successfully in all the patients. Bipedicular injection was used in 2 of the patients, and unipedicular injection was made in the others. The operation time ranged from 25 to 120 minutes (45 minutes per vertebra on average). The average bleeding volume was about 20 ml. Polymethylmethacrylate 1.5-5.0 ml (mean, (3.15+/-0.78) ml) was injected through each pedicle into all the patients except one, who received calcium sulphate 3.5 ml instead. The patients were followed up for 12-15 months (13.5 months on average). The mean visual analogue scale (VAS) score, Oswestry Disability Index, anterior, midline, and posterior vertebral height, and kyphotic angle of the patients were improved significantly at the end of the follow-up compared with those before the operation. (2.5+/-1.3, 35.1%, (20.94+/-6.15) mm, (20.26+/-4.59) mm, (26.72+/-3.49) mm, and 8.2 degrees vs. 8.5+/-1.9, 61.2%, (19.11+/-6.72) mm, (15.88+/-5.73) mm, (25.78+/-3.67) mm, and 17.3 degrees; all P<0.05). The cement distribution with unipedicular injection was mostly limited within the injection site in the vertebral body. Cement extravasation was seen at ten levels (33.3%).</p><p><b>CONCLUSIONS</b>PKP with Sky-bone expander is an effective and relatively safe alternative to the PKP using balloon expander. It can relieve pain, improve physical function, and restore the height of the collapsed vertebrae, but the cement extravasation is unsolved.</p>


Subject(s)
Adult , Aged , Female , Humans , Male , Middle Aged , Bone Cements , Fractures, Compression , General Surgery , Prospective Studies , Spinal Fractures , General Surgery , Spine , General Surgery , Tissue Expansion Devices , Treatment Outcome
2.
Acta Academiae Medicinae Sinicae ; (6): 153-155, 2005.
Article in Chinese | WPRIM | ID: wpr-343749

ABSTRACT

<p><b>OBJECTIVE</b>To test the changes of the stiffness of the intact, released, and instrumented spines in an in vitro porcine model.</p><p><b>METHODS</b>Twelve porcine spines (12 segments each) were harvested for the biomechanical tests with Material Test System. Stiffness during flexion, extension, lateral bending, and axial rotation were recorded; then the specimen was released with intervertebral discs and the facet joints removed, followed by repeating the biomechanical tests for stiffness; and finally, a double-rod titanium construct was applied for internal fixation to each released spine and stiffness tests were repeated again.</p><p><b>RESULTS</b>Compared with the intact porcine spines [stiffness during flexion, extension, lateral bending, and axial rotation was 52.89 +/- 15.98, 105.43 +/- 56.38, 42.09 +/- 14.73, and (16.94 +/- 4.85) N x mm/degrees, respectively], the stiffness of the released porcine spines [stiffness during flexion, extension, lateral bending, and axial rotation was 44.04 +/- 13.73, 41.46 +/- 10.80, 31.75 +/- 7.23, and (9.10 +/- 2.07)N x mm/degrees, respectively] significantly decreased (P < 0.05), while significantly increased stiffness was found in the instrumented porcine spines [stiffness during flexion, extension, lateral bending, and axial rotation was 385.96 +/- 143.25, 138.96 +/- 59.41, 152.56 +/- 87.15, and (55.91 +/- 16.49) N x mm/degrees, respectively] (P < 0.05).</p><p><b>CONCLUSION</b>Higher instant stiffness was found in instrumented spine than the intact one during flexion, extension, lateral bending and axial rotation.</p>


Subject(s)
Animals , Biomechanical Phenomena , Diskectomy , Fracture Fixation, Internal , Internal Fixators , Lumbar Vertebrae , General Surgery , Range of Motion, Articular , Physiology , Spinal Fusion , Methods , Spine , General Surgery , Stress, Mechanical , Swine , Thoracic Vertebrae , General Surgery